Device product
Hamilton-MR1, REF: 161010, Intensive Care Ventilator
Z-2528-2023
Product summary
- Event
- Event 92873
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 12 units
- Official record key
device-enforcement:Z-2528-2023
Official wording
Reason: Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Code information: UDI: 07630002800761 & 07630002813556 / Affected Serial Numbers: 3422 3423 3424 3425 3426 3427 3428 3429 3469 3503 3510 3800
Distribution pattern: Distribution US nationwide, including Puerto Rico. OUS: Not provided
Derived failure modes
-
Unknown
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.