Device product
ICD-VR DVBB1D4 EVERA XT DF4 US, Model Number DVBB1D4; Implantable Cardioverter Defibrillators
Z-1796-2023
Product summary
- Event
- Event 92328
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 4 units
- Official record key
device-enforcement:Z-1796-2023
Official wording
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Code information: GTIN 00643169720435, Lot Serial Number: BWH224619H, BWH224621H, BWH224620H, BWH224623H
Distribution pattern: US Nationwide - Worldwide Distribution
Derived failure modes
-
Unknown
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.