Radiation therapy software, RayStation standalone software treatment planning system. Model numbers: 8B, 8B SP1, 8B SP2, 9A, 9B, 9B SP1, 10A, 10A SP1 and 10B. Software versions: 8.1, 9.0, 9.1, 9.2, 10.0, 10.1.

Z-2039-2023

January 22, 2021

Class II

Product summary

Firm
RaySearch America Inc
Event
Event 92441
Status
Terminated
Classification
Class II
Quantity
5 systems
Official record key
device-enforcement:Z-2039-2023

Official wording

Reason: A software bug affecting results when using deep learning (DL) segmentation followed by geometrical functions: Create wall function, Expand Contract function or ROI algebra function, the results are incorrect. If dose statistics for an incorrectly created ROI are used to determine if a plan is acceptable for treatment, this could potentially lead to an inappropriate plan being approved. The error could lead to a more conservative or more liberal plan than intended.

Code information: UDI-DI: 07350002010129 UDI-DI: 07350002010204 UDI-DI: 07350002010235 UDI-DI: 07350002010174 UDI-DI: 07350002010266 UDI-DI: 07350002010297 UDI-DI: 07350002010303 UDI-DI: 07350002010365 UDI-DI: 07350002010310

Distribution pattern: US: AZ, CA, GA & WA

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software bug