Device product
Radiation therapy software, RayStation standalone software treatment planning system. Model numbers: 8B, 8B SP1, 8B SP2, 9A, 9B, 9B SP1, 10A, 10A SP1 and 10B. Software versions: 8.1, 9.0, 9.1, 9.2, 10.0, 10.1.
Z-2039-2023
Product summary
- Event
- Event 92441
- Status
- Terminated
- Classification
- Class II
- Quantity
- 5 systems
- Official record key
device-enforcement:Z-2039-2023
Official wording
Reason: A software bug affecting results when using deep learning (DL) segmentation followed by geometrical functions: Create wall function, Expand Contract function or ROI algebra function, the results are incorrect. If dose statistics for an incorrectly created ROI are used to determine if a plan is acceptable for treatment, this could potentially lead to an inappropriate plan being approved. The error could lead to a more conservative or more liberal plan than intended.
Code information: UDI-DI: 07350002010129 UDI-DI: 07350002010204 UDI-DI: 07350002010235 UDI-DI: 07350002010174 UDI-DI: 07350002010266 UDI-DI: 07350002010297 UDI-DI: 07350002010303 UDI-DI: 07350002010365 UDI-DI: 07350002010310
Distribution pattern: US: AZ, CA, GA & WA
Derived failure modes
-
Device software or design
software bug