Device product
VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18 slides per cartridge)
Z-0236-2024
Product summary
- Event
- Event 93238
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 342 units
- Official record key
device-enforcement:Z-0236-2024
Official wording
Reason: Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.
Code information: Product Code 1668409: UDI-DI: 10758750009510, Affected Lots: 5905-3348-1557; Product Code 8297749: UDI-DI: 10758750004683, Affected Lots: 5905-3348-2899
Distribution pattern: US States: CA, CO, FL, GA, IL, IN, KS, NC, NY, TN, TX, VT. Canada, France, India, Japan, South Korea
Derived failure modes
-
Unknown
Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.