Device product

Disposable Spetzler-Malis Bipolar forceps, 23CM IM DISP BAYONET 0.5MM TIP, Catalog Number 6770-230-005

Z-0250-2024

September 11, 2023

Class II

Product summary

Firm
Stryker Corporation
Event
Event 93211
Status
Ongoing
Classification
Class II
Quantity
2,321 total units
Official record key
device-enforcement:Z-0250-2024

Official wording

Reason: An error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.

Code information: GTIN: 7613327300161; Lot Numbers: 200515009, 201008013, 201015011, 201126013

Distribution pattern: US Nationwide. Australia, Canada, Japan, Korea, Singapore.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    An error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.