Device product
ICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1; Implantable Cardioverter Defibrillators
Z-1791-2023
Product summary
- Event
- Event 92328
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 71 units
- Official record key
device-enforcement:Z-1791-2023
Official wording
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Code information: GTIN 00643169720756, Lot Serial Numbers: BWR600821S, BWR600947S, BWR600936S, BWR600937S, BWR600941S, BWR600946S, BWR600959S, BWR601046S, BWR600938S, BWR601096S, BWR601093S, BWR600979S, BWR601092S, BWR600997S, BWR600934S, BWR600965S, BWR601050S, BWR601049S, BWR600823S, BWR601053S, BWR600983S, BWR600989S, BWR600984S, BWR601055S, BWR601057S, BWR601098S, BWR600822S, BWR600980S, BWR600981S, BWR600982S, BWR601097S, BWR601100S, BWR601095S, BWR601101S, BWR600991S, BWR601051S, BWR601058S, BWR601099S, BWR601094S; GTIN BWR601149S, BWR601155S, BWR601137S, BWR601138S, BWR601108S, BWR601113S, BWR601111S, BWR601119S, BWR601144S, BWR601146S, BWR601147S, BWR601148S, BWR601067S, BWR601150S, BWR601152S, BWR601115S, BWR601105S, BWR601112S, BWR601059S, BWR601061S, BWR601062S, BWR601063S, BWR601072S, BWR601073S, BWR601078S, BWR601081S, BWR601090S, BWR601116S, BWR601103S, BWR601075S, BWR601102S, BWR601104S
Distribution pattern: US Nationwide - Worldwide Distribution
Derived failure modes
-
Unknown
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.