Recall Observatory FDA recall evidence

Device product

Boston Scientific Model 3120 ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) installed with Model 2892 Programmer Application Software used for the following pacemaker products: ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus. ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus.

Z-0673-2024

November 30, 2023

Class II

Product summary

Firm
Boston Scientific Corporation
Event
Event 93519
Status
Ongoing
Classification
Class II
Quantity
14,328 ZOOM programmers
Official record key
device-enforcement:Z-0673-2024

Official wording

Reason: The software application on the Programmer/Recorder/Monitor (PRM) for Implantable Pulse Generators will display certain diagnostic dates incorrectly, with a year in the 1990s.

Code information: All serial numbers. GTIN numbers in the U.S.: 00802526434723, 00802526456961, 00802526467653, 00802526468582, 00802526473449, 00802526474859, 00802526475788, 00802526477676, 00802526478239, 00802526479663, 00802526480638, 00802526484599, 00802526486906, 00802526494406, 00802526497605, 00802526516009, 00802526523212, 00802526535406, 00802526554209, 00802526562600, 00802526564703, 00802526575303, 00802526579707, 00802526582004, 00802526602900, 00802526604584, 00802526606366, 00802526606403, and 00802526611322. GTIN numbers OUS: 00802526435430, 00802526468582, 00802526479663, 00802526480638, 00802526486906, 00802526478246, 00802526479670, 00802526484605, 00802526486913, 00802526496608, 00802526496615, 00802526506703, 00802526515903, 00802526517112, 00802526526602, 00802526527609, 00802526554803, 00802526561900, 00802526564802, 00802526574801, 00802526579905, 00802526582011, 00802526602917, 00802526605406, 00802526606342, 00802526611308, 00802526615603, 00802526437458, 00802526441974, 00802526444814, 00802526467288, 00802526469343, 00802526475450, 00802526475757, 00802526477669, 00802526478253, 00802526479687, 00802526484612, 00802526486920, 00802526496622, 00802526496639, 00802526515910, 00802526517136, 00802526526800, 00802526527616, 00802526554810, 00802526564819, 00802526574818, 00802526581021, 00802526582028, 00802526605437, 00802526606311, 00802526612428, 00802526615634, 00802526454301, 00802526467295, 00802526469350, 00802526475467, 00802526475764, 00802526477676, 00802526484803, 00802526486081, 00802526486937, 00802526497629, 00802526515927, 00802526532832, 00802526557002, 00802526565427, 00802526582301, 00802526587405, 00802526602924, 00802526606359, 00802526612442, 00802526445262, 00802526454950, 00802526474170, 00802526475474, 00802526475771, 00802526480652, 00802526486944, 00802526497643, 00802526517174, 00802526526817, 00802526527623, 00802526557019, 00802526565304, 00802526579929, 00802526583803, 00802526602931, 00802526605468, 00802526606304, 00802526612411, 00802526615627, 00802526442827, 00802526444807, 00802526469374, 00802526477683, 00802526482267, 00802526497650, 00802526517198, 00802526540400, 00802526555169, 00802526564826, 00802526582325, 00802526585104, 00802526586316, 00802526586316, 00802526603013, 00802526606328, 00802526612435, 00802526615429, 00802526474057, 00802526477522, 00802526479649, 00802526482274, 00802526486470, 00802526497667, 00802526517211, 00802526540509, 00802526555206, 00802526564833, 00802526582318, 00802526585111, 00802526586408, 00802526603006, 00802526605420, 00802526606335, 00802526612404, 00802526615504, 00802526480669, 00802526605451, 00802526496653, 00802526526824, 00802526527630, 00802526527630, 00802526564840, 00802526581038, 00802526582035, 00802526584602, 00802526582035, 00802526605451, 00802526606403, and 00802526612459.

Distribution pattern: Distribution was nationwide, including Guam, Puerto Rico, and the U.S. Virgin Islands. There was also government/military distribution. Foreign distribution was made to Andorra, Argentina, Aruba, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, British Virgin Islands, Bulgaria, Cameroon, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Faroe Islands, Finland, France, French Polynesia, Germany, Great Britain, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Madagascar, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Yemen, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan R.O.C., Thailand, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, Uruguay, Venezuela, Vietnam, and Zimbabwe.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software application on the Programmer/Recorder/Monitor (PRM) for Implantable Pulse Generators will