Device product
Impella CP intravascular micro axial blood pump, Product Number 0048-0032
Z-2115-2023
Product summary
- Firm
- Abiomed, Inc.
- Event
- Event 92321
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 9252 units
- Official record key
device-enforcement:Z-2115-2023
Official wording
Reason: There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.
Code information: UDI-DI: 00813502011388;
Distribution pattern: Worldwide distribution - US Nationwide.
Derived failure modes
-
Unknown
There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.