Alinity ci-series System Control Module, REF: 03R70-01

Z-0613-2024

May 30, 2023

Class II

Product summary

Firm
Abbott Laboratories
Event
Event 93483
Status
Ongoing
Classification
Class II
Quantity
8,424
Official record key
device-enforcement:Z-0613-2024

Official wording

Reason: There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower, including: 1) SCC Reagent load error during daily maintenance; 2) Sample Laboratory Report error; 3) Customer requests mechanism to prevent user error when creating new lots of value assigned Calibrators; 4) Error in processing tests after CCCintController encountered an error with the assay information downloaded from SCC; 5) The assay editor does not correctly check sample dilution total volume limits; 6) User-Applied Labels not adhering to Reagent Bottles; 7) ICT Module did not expire after warranty was exceeded. These performance issues could lead to erroneous results for multiple analytes.

Code information: UDI-DI: 00380740137410. Software Version 3.4.0 and lower.

Distribution pattern: US: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV OUS: ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BERMUDA, BOSNIA AND HERZEGOVINA, BOTSWANA, BRAZIL, BRUNEI, BULGARIA, BURKINA FASO, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FAROE ISLANDS, FINLAND, FRANCE, , GAZA & JERICHO,GERMANY, GREECE, GREENLAND, GUATEMALA, HONDURAS, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOSOVO, KUWAIT, LATVIA, LEBANON, LITHUANIA, MACEDONIA, MADAGASCAR, MALAYSIA, MALDIVES, MAURITANIA, MAURITIUS, MEXICO, MOLDOVA, MONTENEGRO, MOROCCO, NAMIBIA, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NICARAGUA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SENEGAL, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD & TOBAGO, TUNISIA, TURKEY, UAE, UGANDA, UK, UKRAINE, URUGUAY, VIETNAM, ZAMBIA

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software versions 3.4.0 and lower, including: 1) SCC Reagent load error