MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, PASS Tray REF: 8888103004HP

Z-2375-2023

June 28, 2023

Class I

Product summary

Firm
Covidien LP
Event
Event 92660
Status
Ongoing
Classification
Class I
Quantity
271 units
Official record key
device-enforcement:Z-2375-2023

Official wording

Reason: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Code information: GTIN:20884521128047 LOT Numbers: 2133700227

Distribution pattern: US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom

Derived failure modes

  • Foreign material or chemical contamination

    reason.foreign_material_chemical_contamination · v1.0.0

    particulate