Device product
ICD COBALT XT HF QUAD OUS, Model Number DTPA2QQG; Implantable Cardioverter Defibrillators
Z-1780-2023
Product summary
- Event
- Event 92328
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 9 units
- Official record key
device-enforcement:Z-1780-2023
Official wording
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Code information: GTIN 00763000501020, Lot Serial Numbers: RTC614024S, RTC617998S, RTC614026S, RTC617999S, RTC618002S, RTC618007S, RTC621332S, RTC614023S, RTC611661S
Distribution pattern: US Nationwide - Worldwide Distribution
Derived failure modes
-
Unknown
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.