Device product

FreeStyle Libre Reader, REF: 71951-01, 71952-01, 71953-01 a component of the FreeStyle Libre 2 Flash Glucose Monitoring System

Z-1275-2023

February 13, 2023

Class I

Product summary

Firm
Abbott Diabetes Care, Inc.
Event
Event 91756
Status
Ongoing
Classification
Class I
Quantity
1,534,532
Official record key
device-enforcement:Z-1275-2023

Official wording

Reason: Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use.

Code information: REF/UDI-DI/Lot: 71951-01/357599805005/02C313D, 01C256D, 03C222D ,01C174D ,03C166D, 03C152D, 01C143D, 02C115D, 01C084D, 02C087D, 01C073D, 03B251D, 04B251D, 02B251D, 02B109D, 02B082D, 03B082D, 03B014D, 04A350D, 02A289D, 02A279D; 71952-01/357599804008/01B179D, 02B176D, 02B174D, 03B174D, 02B172D, 03B172D, 01B172D, 02B163D, 01B163D, 02B159D, 02B160D, 01B160D, 04B154D, 05B154D, 02B154D, 04B152D, 03B152D, 01B147D, 02B147D, 03B144D, 01B131D, 01B133D, 02B132D, 01B132D, 03B131D, 02B131D, 01B125D, 02B124D, 05B124D, 05B120D, 01B118D, 02B118D, 01B116D, 02B116D, 03B116D, 04B112D, 02B112D, 04B111D, 01B111D, 02B111D, 01B110D, 01B103D, 02B103D, 01B100D, 01B098D, 03B097D, 03B096D, 02B096D, 04B095D, 02B095D, 03B095D, 01B086D, 01B088D, 02B088D, 01B085D, 02B085D, 03B085D, 02B084D, 01B076D, 04B070D, 02B069D, 02B064D, 03B064D, 01B064D, 01B053D, 02B053D, 02B050D, 03B041D, 02B037D, 03B037D, 04B037D, 05B034D, 01B033D, 02B033D, 03B032D, 06B021D, 01B018D, 01B014D, 02B004D, 01A346D, 05A344D, 01A343D, 04A340D, 03A340D, 02A338D, 01A330D, 02A329D, 02A322D, 01A322D, 02A321D, 01A316D, 03A316D, 02A315D, 03A311D, 02A311D, 03A310D, 01A311D, 01A308D, 01A302D, 01A301D, 01A296D, 02A296D, 01A289D, 03A288D, 03A275D, 02A275D, 01A272D, 01A273D, 02A272D, 03A269D, 03A263D, 01A258D, 01A256D, 02A256D, 02A255D, 01A255D, 01A253D, 01A254D, 01A252D, 03A233D, 03A225D, 01A223D, 01A224D, 02A220D, 03A220D, 01A220D, 01A210D, 02A210D, 02A209D, 01A206D, 02A206D, 03A206D, 04A205D; 71953-01/357599803001/06B074F, 01C346D, 03C340D, 01C341D, 02C340D, 01C340D, 02C334D, 04C334D, 01C334D, 01C327D, 01C332D, 01C326D, 01C325D, 02C325D, 01C322D, 02C321D, 03C320D, 01C318D, 01C317D, 01C313D, 02C312D, 01C307D, 01C308D, 03C305D, 01C305D, 02C305D, 01C300D, 02C298D, 01C299D, 01C298D, 03C294D, 06C292D, 02C294D, 04C292D, 02C287D, 01C283D, 03C279D, 02C278D, 01C273D, 04C276D, 01C272D, 01C276D, 08C271D, 01C266D, 03C270D, 02C265D, 02C262D, 02C257D, 03C257D, 01C257D, 02C256D, 03C256D, 02C255D, 03C251D, 02C250D, 01C249D, 02C245D, 01C245D, 04C243D, 02C242D, 02C243D, 01C242D, 01C243D, 03C242D, 04C241D, 02C237D, 03C238D, 01C237D, 03C236D, 02C236D, 01C228D, 02C229D, 03C224D, 01C227D, 01C224D, 02C224D, 01C223D, 02C223D, 03C213D, 01C207D, 01C209D, 02C208D, 03C207D, 01C201D, 02C199D, 05C199D, 03C199D, 03C196D, 02C193D, 01C194D, 01C188D, 02C188D, 01C186D, 01C182D, 01C181D, 02C181D, 02C172D, 05C165D, 03C165D, 01C161D, 01C159D, 03C159D, 01C146D, 05C146D, 01C144D, 02C145D, 02C143D, 01C140D, 02C140D, 03C140D, 04C140D, 04C138D, 01C137D, 01C138D, 01C133D, 01C136D, 02C136D, 03C133D, 02C133D, 02C130D, 03C129D, 03C126D, 03C123D, 03C117D, 02C116D, 02C112D, 01C109D, 02C080D, 01C075D, 01C077D, 02C074D, 03C075D, 02C073D, 03C074D, 01C074D, 01C034F, 02C032F, 01C028F, 02C031F, 03C031F, 03C028F, 02C025F, 02C013D, 01C014D, 03C013D, 01C011D, 01C012D, 02C010D, 02C005D, 01C006D, 03C005D, 02B363D, 01B362D, 02B362D, 01B356D, 04B356D, 05B356D, 02B356D, 04B354D, 03B351D, 04B352D, 06B349F, 02B344F, 01B343F, 04B302F, 01B342F, 02B342F, 02B343F, 01B341F, 02B341F, 01B341D, 02B341D, 01B302F, 01B340F, 02B340F, 03B302F, 04B301F, 01B337D, 01B337F, 02B337F, 01B336F, 02B336F, 01B335F, 03B335F, 02B333D, 01B334D, 01B333D, 01B308D, 02B328F, 01B327F, 03B326F, 04B326F, 01B323F, 01B326F, 06B322F, 01B305D, 07B322F, 05B322F, 01B313D, 03B312D, 03B313D, 02B312D, 01B310D, 01B308F, 01B309F, 01B309D, 02B308F, 03B306F, 01B306F, 02B306F, 01B306D, 02B305D, 01B289D, 02B289D, 01B288D, 02B229D, 02B216D, 01B215D, 02B211D, 02B208D, 01B208D, 01B205D, 01B207D, 03B203D, 02B203D, 03B109D, 05B082D, 02B083D, 02B082F, 04B082F, 02B081F, 03B081F, 02B078F, 05B078F, 01B077F, 05B076F, 02B076F, 05B075F, 04B074F, 05B071F, 01B071F, 02B046F, 05B043F, 06B046F, 09B046F, 02B042F, 03B043F, 05B046F, 03B041F, 05B041F, 03B036F, 01B036F, 01B039F, 03B039F, 02B036F, 05B035F, 02B035F, 04B034F, 04B035F, 06B034F, 07B034F, 02B034F, 11B021F, 02B013F, 04B013F, 01B015F, 01B012F, 01B014F, 05B013F, 03B012F, 01B008F, 01B011F, 01B007F, 02B007F, 01A364D, 02A363D, 02A358D, 01A352D, 03A352D, 02A351D, 03A350D, 03A344D, 02A340D, 01A328D, 03A325D, 04A325D, 02A304D, 02A303D, 02A297D, 02A298D, 03A296D, 01A290D, 02A283D, 01A283D, 01A281D, 02A280D, 03A268D, 01A267D, 02A268D, 03A266D, 04A262D, 01A259D, 01A260D, 02A248D, 02A240D, 06A232D, 02A232D, 01A231D, 03A231D, 04A225D, 05A225D, 01A219D, 04A218D, 05A218D, 06A218D, 02A212D, 01A211D, 02A211D, 03A210D, 02A205D, 03A205D, 01A204D, 02A203D, 04A202D, 01A202D, 04A198D, 05A198D, 01A198D, 02A198D, 03A198D

Distribution pattern: U.S. Nationwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use.