CRT-D DTBB2Q1 VIVA QUAD S IS4/DF1 INTL, Model Number DTBB2Q1; Implantable Cardioverter Defibrillators

Z-1747-2023

May 10, 2023

Class I

Product summary

Firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Event
Event 92328
Status
Ongoing
Classification
Class I
Quantity
7 units
Official record key
device-enforcement:Z-1747-2023

Official wording

Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Code information: GTIN 00643169980181, Lot Serial Numbers: BLQ600039S, BLQ600033S, BLQ600034S, BLQ600035S, BLQ600036S, BLQ600037S, BLQ600038S

Distribution pattern: US Nationwide - Worldwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.