Device product

Single Use Repositionable Clip QUICKCLIP PRO 165CM 5/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202LR.B

Z-1988-2023

May 04, 2023

Class II

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 92325
Status
Ongoing
Classification
Class II
Quantity
45 pieces
Official record key
device-enforcement:Z-1988-2023

Official wording

Reason: Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.

Code information: UDI-DI: 04953170385926 Lot Numbers: 24K, 26K, 27K, 29K, 2XK, 2YK

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.