Skip to content
Recall Observatory FDA recall evidence

Device product

DeVilbiss, 525 5-liter Oxygen Concentrator Models 525DS, 525DS-Q

Z-0724-2023

November 18, 2022

Class II

Product summary

Firm
DeVilbiss Healthcare LLC
Event
Event 91112
Status
Ongoing
Classification
Class II
Quantity
1004315 (US); 414559 (OUS)
Official record key
device-enforcement:Z-0724-2023

Official wording

Reason: The firm has received complaints of the device becoming hot. The firm has updated the user manual to include warnings that applied parts in contact with the patient may in some situations exceed 41 degrees Celsius (105.8 degrees Fahrenheit), clarification on expected service life, and the need for maintenance to certain components.

Code information: Model 525DS UDI-DI 00885304000846 All Lots Model 525DS-Q UDI-DI 00885304009689 All Lots

Distribution pattern: US Nationwide. Foreign distribution pending.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The firm has received complaints of the device becoming hot. The firm has updated the user manual to include warnings that applied parts in contact with the patient may in some situations exceed 41 degrees Celsius (105.8 degrees Fahrenheit), clarification on expected service life, and the need for maintenance to certain components.