Skip to content
Recall Observatory FDA recall evidence

Device product

a. Skin Adhesive McKesson LiquiBand Exceed 0.4 mL Liquid Dome Applicator Tip 2-Octyl Cyanoacrylate. Model Number: 122-LBXS. b. Skin Adhesive McKesson LiquiBand Exceed 0.8 mL Liquid Precision and Dome Applicator Tip 2-Octyl Cyanoacrylate. Ref: 122-LBX. c. ADVANCED MEDICAL SOLUTIONS McKesson ADHESIVE, LIQUIBAND TISSUE FLOW SM TIP 0.5ML (12/BX 6BX/CS). Catalog #122-LFC

Z-1788-2022

May 25, 2022

Class III

Product summary

Firm
Mckesson Medical-Surgical Inc. Corporate Office
Event
Event 90731
Status
Ongoing
Classification
Class III
Quantity
a. 46.9 bx b. 75.1 bx c. 13.4 bx
Official record key
device-enforcement:Z-1788-2022

Official wording

Reason: Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. This device with this specific storage temperature deviation is not likely to cause adverse health consequences.

Code information: a. UDI-DI :10612479232256; 40612479232271; 20612479232260. Lot codes distributed between July and September 2021. b. GTIN: 10612479208183; 20612479195978; 40612479195989. c. GTIN: 20612479185863; 10612479208206; 40612479185874; 20612479185863

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. This device with this specific storage temperature deviation is not likely to cause adverse health consequences.