HeartWare HVAD Pump Kit, REF 1104CA-CLIN
Z-0747-2023
Product summary
- Firm
- Heartware, Inc.
- Event
- Event 91281
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 6 units
- Official record key
device-enforcement:Z-0747-2023
Official wording
Reason: Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop, if an urgent driveline connection or controller issue needs to be addressed. In the event a hardened driveline cover is observed, contact your Medtronic field representative for further assessment and/or servicing.
Code information: Serial Numbers: HW24101, HW24102, HW24103, HW24104, HW24755, HW24763
Distribution pattern: Worldwide distribution.
Derived failure modes
-
Unknown
Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop, if an urgent driveline connection or controller issue needs to be addressed. In the event a hardened driveline cover is observed, contact your Medtronic field representative for further assessment and/or servicing.