Device product
BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150, BF-1T60, BF-P150 BF-P60 and EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number XT160
Z-0189-2024
Product summary
- Event
- Event 93052
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 15942 devices
- Official record key
device-enforcement:Z-0189-2024
Official wording
Reason: There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
Code information: Model BF-1T150 UDI-DI: 4953170308185; Model BF-1T60 UDI-DI: 4953170339264; Model BF-P150 UDI-DI: N/A; Model BF-P60 UDI-DI: 4953170339196; Model BF-XT160 UDI-DI: N/A
Distribution pattern: US Nationwide distribution
Derived failure modes
-
Unknown
There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.