Skip to content
Recall Observatory FDA recall evidence

Device product

Sterile Procedural Trays, labeled as the following: a. ST. ANNE'S MAJOR BASIN SET UP b. SINGLE BASIN SET c. DOUBLE BASIN SET d. LAPAROTOMY BASIN e. ORTHOPEDIC CUSTOM BASIN SET f. OISC BASIN TRAY g. BASIN PACK h. BASIN SET i. SURGI START PACK j. PK, GEN-SET UP - MIDWEST k. SINGLE BASIN PACK-LF l. AGHM MINOR SINGLE PACK m. MINOR PACK n. SINGLE BASIN PACK o. MAJOR BASIN PACK p. BASIC SET UP-LF q. BASIC SET-UP-LF r. BASIN PACK s. MINOR BASIN t. AMC MAJOR BASIN u. AMC MAJOR BASIN

Z-1078-2023

November 17, 2022

Class II

Product summary

Firm
MEDLINE INDUSTRIES, LP - Northfield
Event
Event 91386
Status
Ongoing
Classification
Class II
Quantity
2.542 cases (6,739 units)
Official record key
device-enforcement:Z-1078-2023

Official wording

Reason: Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.

Code information: a. ST. ANNE'S MAJOR BASIN SET UP, Item Number: DYNJ24628M, Case UDI/GTIN: 40195327109135, Unit UDI/GTIN: 10195327109134, Lot Number: 22JLA323,22ILA590; b. SINGLE BASIN SET, Item Number: DYNJ35592K, Case UDI/GTIN: 40195327110827, Unit UDI/GTIN: 10195327110826, Lot Number: 22JBW819,22IBK206; c. DOUBLE BASIN SET, Item Number: DYNJ35596G, Case UDI/GTIN: 40195327110810, Unit UDI/GTIN: 10195327110819, Lot Number: 22JBH675,22HBY465; d. LAPAROTOMY BASIN, Item Number: DYNJ41354B, Case UDI/GTIN: 40195327246311, Unit UDI/GTIN: 10195327246310, Lot Number: 22JBE052; e. ORTHOPEDIC CUSTOM BASIN SET, Item Number: DYNJ47991J, Case UDI/GTIN: 40195327230600, Unit UDI/GTIN: 10195327230609, Lot Number: 22JBX225,22JBH710,22HBW917,22HBV171; f. OISC BASIN TRAY, Item Number: DYNJ48354I, Case UDI/GTIN: 40195327154302, Unit UDI/GTIN: 10195327154301, Lot Number: 22IMF862; g. BASIN PACK, Item Number: DYNJ50123D, Case UDI/GTIN: 40195327201990, Unit UDI/GTIN: 10195327201999, Lot Number: 22HBU014; h. BASIN SET, Item Number: DYNJ50906F, Case UDI/GTIN: 40195327140145, Unit UDI/GTIN: 10195327140144, Lot Number: 22JMB641,22HMD087; i. SURGI START PACK, Item Number: DYNJ52541D, Case UDI/GTIN: 40195327164509, Unit UDI/GTIN: 10195327164508, Lot Number: 22JBH137; j. PK, GEN-SET UP - MIDWEST, Item Number: DYNJ57608B, Case UDI/GTIN: 40195327253296, Unit UDI/GTIN: 10195327253295, Lot Number: 22JMB780; k. SINGLE BASIN PACK-LF, Item Number: DYNJ58389D, Case UDI/GTIN: 40195327144198, Unit UDI/GTIN: 10195327144197, Lot Number: 22GBW998; l. AGHM MINOR SINGLE PACK, Item Number: DYNJ62766B, Case UDI/GTIN: 40195327116775, Unit UDI/GTIN: 10195327116774, Lot Number: 22JMC641,22HMD203; m. MINOR PACK, Item Number: DYNJ63961B, Case UDI/GTIN: 40195327146970, Unit UDI/GTIN: 10195327146979, Lot Number: 22IDA519; n. SINGLE BASIN PACK, Item Number: DYNJ66012A, Case UDI/GTIN: 40195327221455, Unit UDI/GTIN: 10195327221454, Lot Number: 22JLA612,22ILA886,22HLA208; o. MAJOR BASIN PACK, Item Number: DYNJ69406A, Case UDI/GTIN: 40195327112333, Unit UDI/GTIN: 10195327112332, Lot Number: 22JMH995,22JMA882; p. BASIC SET UP-LF, Item Number: DYNJ81646, Case UDI/GTIN: 40195327130269, Unit UDI/GTIN: 10195327130268, Lot Number: 22IBH244; q. BASIC SET-UP-LF, Item Number: DYNJ82390, Case UDI/GTIN: 40195327189540, Unit UDI/GTIN: 10195327189549, Lot Number: 22JBV631,22IBC646; r. BASIN PACK, Item Number: DYNJ82969, Case UDI/GTIN: 40195327218325, Unit UDI/GTIN: 10195327218324, Lot Number: 22JBN923; s. MINOR BASIN, Item Number: DYNJ907808B, Case UDI/GTIN: 40195327244430, Unit UDI/GTIN: 10195327244439, Lot Number: 22JBU478,22JBA257,22IBV340,22IBS662; t. AMC MAJOR BASIN, Item Number: GSP158AA, Case UDI/GTIN: 40195327143511, Unit UDI/GTIN: 10195327143510, Lot Number: 22HBT951; u. AMC MAJOR BASIN, Item Number: GSP158AB, Case UDI/GTIN: 40195327252022, Unit UDI/GTIN: 10195327252021, Lot Number: 22JBD534

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.