Device product

Simplexa COVID-19 Direct, REF: MOL4150; including Reaction Mix, REF: MOL4150; with previous assay definition v2

Z-1209-2023

January 30, 2023

Class II

Product summary

Firm
DiaSorin Molecular LLC
Event
Event 91504
Status
Ongoing
Classification
Class II
Quantity
90142
Official record key
device-enforcement:Z-1209-2023

Official wording

Reason: Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes.

Code information: UDI-DI: 20816101027020. Product REF/Lot/Reaction Mix REF/Lot: MOL4150/US13950/MOL4151/US14091, MOL4150/US14913/MOL4151/US14063, MOL4150/US13952/MOL4151/US14065, MOL4150/US13955/MOL4151/US14086, MOL4150/US13959/MOL4151/US14088, MOL4150/US13956/MOL4151/US14087, MOL4150/US13960/MOL4151/US14093, MOL4150/US13961/MOL4151/US14094, MOL4150/US13973/MOL4151/US13732, MOL4150/US13974/MOL4151/US13736, MOL4150/US14006/MOL4151/US13738, MOL4150/US14004/MOL4151/US15064, MOL4150/US14005/MOL4151/US13737, MOL4150/US14008/MOL4151/US14100, MOL4150/US14007/MOL4151/US14099, MOL4150/US14010/MOL4151/US14102, MOL4150/US14011/MOL4151/US14103, MOL4150/US14012/MOL4151/US15290, MOL4150/US14018/MOL4151/US14111, MOL4150/US14009/MOL4151/US14101, MOL4150/US14017/MOL4151/US14109, MOL4150/15061N/MOL4151/13474N, MOL4150/X15475N/MOL4151/X15479N, MOL4150/13330N/MOL4151/14829N, MOL4150/13331N/MOL4151/13751N, MOL4150/13336N/MOL4151/13752N, MOL4150/13968N/MOL4151/14036N, MOL4150/14346N/MOL4151/14038N, MOL4150/13967N/MOL4151/15082N

Distribution pattern: US Nationwide distribution including in the states of NJ, ME, ND, IL, MI, FL, GA, AZ, MA, CA, MO, SD, MN, PA, OH, VT, NE, WI, CO, DC, VA, TX, NH, IN, IA, MD, NC, AL, NY, KY, WV, RI, KS, NM, OK, WA, SC, CT, TN, NV, LA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes.