Device product
Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN)RFPC-35-180 ORDER NUMBER (GPN) G07937. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
Z-1706-2022
Product summary
- Event
- Event 90710
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 496 units
- Official record key
device-enforcement:Z-1706-2022
Official wording
Reason: Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
Code information: UDI-DI: (01)00827002079373- Lot Numbers: 14145420, 14145424, 14156909, 14156912, 14156916, 14164348, 14173577
Distribution pattern: Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA
Derived failure modes
-
Unknown
Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.