CereLink ICP Monitor; Catalog No. 826820, 826820P. Intracranial pressure monitor.

Z-1395-2022

June 22, 2022

Class I

Product summary

Firm
Integra LifeSciences Corp.
Event
Event 90457
Status
Ongoing
Classification
Class I
Quantity
1,210 units
Official record key
device-enforcement:Z-1395-2022

Official wording

Reason: Integra has received complaints associated with ICP readings drifting to -50 mmHg (out-of-range) and may manifest an error message "sensor or extension cable failure!" The firm has found that this is caused by electrical interference from a component of the monitor's circuit board and from the environment.

Code information: UDI-DI- 10381780533778; All serial numbers.

Distribution pattern: Worldwide Distribution: US nationwide, Australia, New Zealand, Canada, Hong Kong, Pakistan, Georgia, Saudi Arabia, Israel, South Africa, United Arab Emirates, Austria, Belgium, Czech Republic, Denmark, Estonia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Latvia, Lithuania, Netherlands, Norway, Portugal, Switzerland, Spain, Sweden, Slovakia, and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Integra has received complaints associated with ICP readings drifting to -50 mmHg (out-of-range) and may manifest an error message "sensor or extension cable failure!" The firm has found that this is caused by electrical interference from a component of the monitor's circuit board and from the environment.