Device product

INOmax DSIR -(delivery system) delivers INOMAX (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, Model Number: 10007

Z-1745-2022

June 23, 2022

Class II

Product summary

Firm
Mallinckrodt, LLC.
Event
Event 90723
Status
Ongoing
Classification
Class II
Quantity
4,225 units OUS
Official record key
device-enforcement:Z-1745-2022

Official wording

Reason: The INOmax DSIR Delivery System, manufactured by Mallinckrodt Pharmaceuticals, is not compatible with the upgraded version of the Leoni Plus ventilator, manufactured by Lowenstein Medical. The Leoni Plus ventilator was recently upgraded in 2021 by Lowenstein Medical, with new hardware and software, and the upgraded version cannot be appropriately connected with the INOmax DSIR as instructed in the INOmax DSIR Operators Manual (OM). The INOmax DSIR OM requires the use of a one-way valve with the Leoni Plus ventilator during High Frequency Oscillatory Ventilation mode and warns that use of the device without a one-way valve may result in high Nitric Oxide (NO) delivery.

Code information: All serial numbers

Distribution pattern: Foreign Only: Argentina Australia Austria Beligum Canada Chile Columbia Czech Republic Denmark Ecuador Estonia Finland France Germany Great Britain Hungary Iceland Ireland Italy Latvia Mexico Netherlands Norway Poland Portugal Romania Slovakia South Africa Spain Sweden Switzerland UAE Uruguay

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The INOmax DSIR Delivery System, manufactured by Mallinckrodt Pharmaceuticals, is not compatible with the upgraded version of the Leoni Plus ventilator, manufactured by Lowenstein Medical. The Leoni Plus ventilator was recently upgraded in 2021 by Lowenstein Medical, with new hardware and software, and the upgraded version cannot be appropriately connected with the INOmax DSIR as instructed in the INOmax DSIR Operators Manual (OM). The INOmax DSIR OM requires the use of a one-way valve with the Leoni Plus ventilator during High Frequency Oscillatory Ventilation mode and warns that use of the device without a one-way valve may result in high Nitric Oxide (NO) delivery.