NaviCare Nurse Call/Voalte Nurse Call
Z-1306-2022
Product summary
- Event
- Event 90322
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 283 installations
- Official record key
device-enforcement:Z-1306-2022
Official wording
Reason: An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.
Code information: software versions: 3.9.100, 3.9.101,3.9.102, 3.9.200, 3.9.201, and 3.9.300
Distribution pattern: nationwide and Canada
Derived failure modes
-
Unknown
An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.