Skip to content
Recall Observatory FDA recall evidence

Device product

Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN):RLPC-35-180 ORDER NUMBER (GPN) G07516. For vascular catheter positioning and exchange in diagnostic and interventional procedures.

Z-1708-2022

July 19, 2022

Class II

Product summary

Firm
Cook Incorporated
Event
Event 90710
Status
Ongoing
Classification
Class II
Quantity
1450 units
Official record key
device-enforcement:Z-1708-2022

Official wording

Reason: Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

Code information: UDI-DI: (01)00827002075160- Lot Numbers: 14103953, 14103956, 14143728, 14145412, 14145419, 14153927, 14156907, 14156908, 14156914, 14160996

Distribution pattern: Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.