Device product
Gallant DR Implantable Cardioverter Defibrillator REF # CDDRA500Q
Z-0988-2022
Product summary
- Event
- Event 89825
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 1
- Official record key
device-enforcement:Z-0988-2022
Official wording
Reason: Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.
Code information: REF# CDDRA500Q SN# 111018237, Exp Date: 31-Oct-2022 Material# 600115496 UDI# 05415067032003
Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of FL, KY and the country of France, Czech Republic.
Derived failure modes
-
Unknown
Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.