Device product
Artis icono biplane -angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 11327600
Z-1245-2022
Product summary
- Event
- Event 90184
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 51 units
- Official record key
device-enforcement:Z-1245-2022
Official wording
Reason: if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm
Code information: UDI-DI: 04056869063317 Serial Number: 180062 180054 180019 180065 180056 180057 180041 180317 180034 180072 180060 180070 180048 180083 180315 180080 180301 180305 180067 180330 180040 180333 180029 180336 180316 180084 180334 180018 180046 180047 180335 180085 180035 180081 180331 180021 180321 180309 180323 180329 180043 180049 180077 180086 180053 180055 180015 180014 180066 180082 180037
Distribution pattern: US Nationwide distribution.
Derived failure modes
-
Unknown
if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm