Device product

Field replacement back up batteries distributed on or after April 1, 2019 for Avance CS2, Avance CS2 Pro, Avance, Amingo, Aespire View anesthesia devices

Z-1286-2022

April 25, 2022

Class II

Product summary

Firm
GE Healthcare, LLC
Event
Event 90137
Status
Ongoing
Classification
Class II
Quantity
88,347 batteries
Official record key
device-enforcement:Z-1286-2022

Official wording

Reason: Backup batteries can fail earlier than their estimated life.

Code information: Field Replacement Units (FRU) Battery kits - PN: 1009-5682-000-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V) or PN: 5856787-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V PAIR), that may have been installed by the customer into any of the following anesthesia systems: o Avance 1009-9002-000 o Aespire View 1009-9212-000 o Avance CS2 and Avance CS2 Pro 1009-9050-000

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, China, Columbia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Italy, Japan, Kazakhstan, Kenya, Korea, Kuwait, Lao PDR, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Nepal, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Backup batteries can fail earlier than their estimated life.