Device product
Western Oxytote DTE Digital VIPR (head only) Model: MNDR-600
Z-0268-2022
Product summary
- Event
- Event 88877
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 4974 units
- Official record key
device-enforcement:Z-0268-2022
Official wording
Reason: VIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and smoke
Code information: Units distributed from 05/29/20 to 08/31/21. Digital VIPR units manufactured and distributed after 08/31/21 are outside the scope of this recall and will be recognized by a green and silver control knob sticker (green logo, silver background)
Distribution pattern: CT, MD, MT, OH
Derived failure modes
-
Unknown
VIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and smoke