Skip to content
Recall Observatory FDA recall evidence

Device product

BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port equipped with a threaded female luer during blood or urine sample collection using evacuated BD Vacutainer Blood or Urine Collection Tubes Catalog Number: 364902, 36490200

Z-0665-2022

January 06, 2022

Class III

Product summary

Firm
Becton Dickinson & Company
Event
Event 89486
Status
Ongoing
Classification
Class III
Quantity
288,152,790 Units
Official record key
device-enforcement:Z-0665-2022

Official wording

Reason: May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts

Code information: All Lots within Expiry Catalog No: 364902 UDI: (01)50382903649025 Catalog Number: 36490200 UDI:(01)60382903649022

Distribution pattern: US Nationwide distribution in the state of IL.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts