Device product
Olympus URETERO-RENO FIBERSCOPE Model: URF-P6
Z-1263-2022
Product summary
- Event
- Event 90195
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 2 units
- Official record key
device-enforcement:Z-1263-2022
Official wording
Reason: Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.
Code information: UDI-DI: 04953170340802 Serial Number: 2836210 2834747I
Distribution pattern: CA FL MA MD MI NJ OH PR WI
Derived failure modes
-
Unknown
Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.