Skip to content
Recall Observatory FDA recall evidence

Device product

VIDAS DEX@, Dimer Exclusion II, REF 30455-01

Z-0427-2022

September 22, 2021

Class II

Product summary

Firm
bioMerieux, Inc.
Event
Event 88797
Status
Ongoing
Classification
Class II
Quantity
1579
Official record key
device-enforcement:Z-0427-2022

Official wording

Reason: bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to potential delayed results as the user must run another test.

Code information: UDI 03573026387006, Lot #s: 1008696660,1008761020, 1008864290, 1008891690

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to potential delayed results as the user must run another test.