Artis zee floor MN, Model Number 10094142
Z-0971-2022
Product summary
- Event
- Event 89860
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 24 systems
- Official record key
device-enforcement:Z-0971-2022
Official wording
Reason: If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Code information: UDI: 4056869010090; Serial Numbers: 140421 140115 140101 140205 140213 140204 140209 140219 140224 140218 140226 140221 140220 140206 140105 140423 140406 140422 140102 140214 140109 140202 140424 140114
Distribution pattern: US Nationwide.
Derived failure modes
-
Unknown
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".