Skip to content
Recall Observatory FDA recall evidence

Device product

EDI Novel Coronavirus COVID-19 IgG ELISA Kit, REF: KT-1032 Per label: This kit is intended for qualitative detection of human anti-COVID-19 IgG antibody in human serum. This kit is to be used as an aid for the detection of novel COVID-19. Patients with suspected clustering cases require diagnosis or differential diagnosis of novel coronavirus infection.

Z-0426-2021

October 01, 2020

Class II

Product summary

Firm
Epitope Diagnostics, Inc.
Event
Event 86516
Status
Ongoing
Classification
Class II
Quantity
2886
Official record key
device-enforcement:Z-0426-2021

Official wording

Reason: The use of deionized water rather than distilled water during the manufacturing process, may have resulted in accelerated decreased OD signals for the Positive Control value.

Code information: Lot Numbers: P826, P815, P796, P776

Distribution pattern: U.S Distribution.: NH, CO, OR, CA, FL, NY, MD, NJ, IL, NC, MI, TX, PA, MA, OH, NM, VA, SC, HI, and TN.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The use of deionized water rather than distilled water during the manufacturing process, may have resulted in accelerated decreased OD signals for the Positive Control value.