Device product

Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition

Z-0242-2022

June 30, 2021

Class II

Product summary

Firm
GE Healthcare, LLC
Event
Event 89027
Status
Ongoing
Classification
Class II
Quantity
105 in total
Official record key
device-enforcement:Z-0242-2022

Official wording

Reason: The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.

Code information: Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition

Distribution pattern: Worldwide and US Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.