Skip to content
Recall Observatory FDA recall evidence

Device product

SpaceOAR Vue System, UPN SV-1010 and SV-2101. Used to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer to reduce the radiation dose delivered to the anterior rectum.

Z-1003-2022

February 24, 2022

Class II

Product summary

Firm
Boston Scientific Corporation
Event
Event 89862
Status
Ongoing
Classification
Class II
Quantity
23,368 units
Official record key
device-enforcement:Z-1003-2022

Official wording

Reason: Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.

Code information: UPN SV-1010, GTIN 00850009803009, Batch Numbers: 26383209, 26383210, 26517604, 26861584, 26861585, 26891612, 26891613, 27083784, 27190242, 27210560, 27257801, 27257802, 27455966, 27455967, 27496959, 27497140, 27670956, 27708041, 27708042, 27708043, 27930863, 27930864, 27941129, 27941130, 27949315, 28094077, 28247077, 28247078, 28358778, 28377018, 28377019, 28464794, 28464795, 28496461, 28496462, 28578758, 28578759, 28608231, 28608232; UPN SV-2101, GTIN 00864661000140, Batch Numbers: 26345481, 26376961, 26376962, 26376963, 26376964, 26380456, 26380457, 26384055, 26384056, 26408849, 26485974, 26485975, 26485977, 26486586, 26486587, 26486588, 26493461, 26517606, 26517607, 26529719, 26567283, 26567284, 26567285, 26567286, 26567287, 26574373, 26574374, 26574378, 26574379, 26574441, 26611593, 26611594, 26611595, 26652286, 26652287, 26691824, 26691826, 26691827, 26691828, 26705563, 26705564, 26705565, 26705566, 26728564, 26728565, 26743769, 26743770, 26743771, 26743772, 26745057, 26770161, 26770162, 26770163, 26770164, 26770165, 26861586, 26880769, 26880770, 26880771, 26880772, 26898042, 26898043, 26898044, 26898045, 26899812, 26899813, 26906603, 26906604, 26929712, 26929713, 26929714, 26929715, 26929716, 26929717, 26929719, 26929860, 26929861, 26957514, 26957515, 26957516, 26957517, 26969537, 26969538, 26969539, 26970440, 26970441, 26970442, 27010832, 27040235, 27040236, 27060422, 27060423, 27060424, 27060425, 27060426, 27078872, 27078873, 27084457, 27084458, 27084459, 27084960, 27100768, 27100769, 27100770, 27123402, 27123404, 27140485, 27140486, 27163487, 27163488, 27180605, 27190243, 27190244, 27190245, 27190686, 27193837, 27193838, 27193839, 27194460, 27194461, 27197622, 27197623, 27214381, 27214382, 27214383, 27224419, 27257804, 27257805, 27259062, 27259063, 27259064, 27259065, 27268186, 27268187, 27283462, 27283463, 27283464, 27283465, 27283466, 27317224, 27317225, 27317226, 27329273, 27329274, 27329275, 27358004, 27358005, 27361380, 27361381, 27382203, 27382204, 27416003, 27416004, 27416005, 27416006, 27416007, 27444898, 27444899, 27445360, 27447905, 27469811, 27469812, 27469813, 27475230, 27496957, 27496958, 27512052, 27512053, 27512054, 27512055, 27517685, 27518868, 27518869, 27528492, 27531152, 27531153, 27531154, 27531155, 27531156, 27531157, 27601494, 27601495, 27657444, 27657445, 27657446, 27670942, 27670945, 27670946, 27670949, 27705891, 27720117, 27720118, 27720119, 27745215, 27745216, 27756475, 27769338, 27769339, 27769860, 27769861, 27769862, 27769863, 27797083, 27797084, 27797085, 27797086, 27797087, 28094078, 28094079, 28094320, 28103330, 28104177, 28104178, 28104179, 28104420, 28110684, 28110685, 28110686, 28110687, 28116634, 28116635, 28138219, 28138440, 28146630, 28146631, 28169753, 28169754, 28169755, 28179669, 28179670, 28179671, 28202160, 28237387, 28237388, 28237389, 28247079, 28247160, 28247161, 28247162, 28247269, 28247270, 28247271, 28247272, 28254386, 28254387, 28254388, 28265943, 28265944, 28265945, 28265946, 28265947, 28272556, 28272557, 28272558, 28272559, 28272560, 28287355, 28287356, 28294718, 28294719, 28311111, 28311112, 28323414, 28323415, 28323416, 28323417, 28335613, 28335614, 28350366, 28350367, 28350368, 28350369, 28364006, 28364007, 28364008, 28371227, 28371228, 28371229, 28389635, 28389636, 28412353, 28419102, 28419103, 28432892, 28432893, 28473289, 28473290, 28493777, 28493778, 28499505, 28499506, 28499507, 28555843, 28565355, 28565356, 28565357, 28587389, 28587390, 28587391, 28587392, 28587393, 28587394, 28605867, 28606252, 28611099, 28612440, 28612441, 28612442, 28613189, 28635760, 28635761, 28635762, 28635763, 28654059, 28654440, 28654441, 28663263, 28663264, 28663265, 28668946, 28668947, 28668948, 28683114, 28683115, 28686271, 28686272, 28686273, 28694687, 28707785, 28709642, 28709643, 28714572, 28723144, 28734763, 28734764, 28758266, 28766992, 28766993, 28766994, 28766995, 28766996, 28811555, 28811556, 28811558, 28811560, 28819972, 28819973, 28821692, 28821693, 28822526, 28822527, 28828281, 28828282, 28836638, 28836639, 28838125

Distribution pattern: Worldwide - US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.