VITROS 4600 Chemistry System, Version 3.3.1 & below For use in the in vitro quantitative measurement of a variety of analytes of clinical interest in certain biological fluids (depending on the analyte) such as serum, plasma, urine and cerebrospinal fluid.

Z-0653-2019

October 19, 2018

Class II

Product summary

Firm
Ortho-Clinical Diagnostics
Event
Event 81563
Status
Terminated
Classification
Class II
Quantity
728
Official record key
device-enforcement:Z-0653-2019

Official wording

Reason: Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.

Code information: Product Code: 6802445 UDI: 10758750012343 Serial Numbers: J46000108-J46001132

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software anomaly