Skip to content
Recall Observatory FDA recall evidence

Device product

Beaver Xstar Crescent Knife, 2.5mm, 55 degrees, bevel down. Intended for or cataract surgery. REF#: 373809

Z-0227-2022

October 08, 2021

Class II

Product summary

Firm
Beaver Visitec International, Inc.
Event
Event 88826
Status
Ongoing
Classification
Class II
Quantity
5950 eaches
Official record key
device-enforcement:Z-0227-2022

Official wording

Reason: Product packaged as Beaver Xstar Crescent Knife, 2.5mm, 55 degrees bevel-down blades contain bevel-up blades

Code information: LOT#: 6000842, 6002045, 6011609, 6034238, 6040517, 6042907. UDI: " Primary: 00886158000594 " Secondary: 30886158000595 " Tertiary: 50886158000599

Distribution pattern: Nationwide Foreign: Australia Austria France Germany Netherlands Spain Japan Russia South Africa UK

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Product packaged as Beaver Xstar Crescent Knife, 2.5mm, 55 degrees bevel-down blades contain bevel-up blades