Device product

Phantom Nail System, Drill, 4.6 x 300mm, Cannulated, 3/16" Sq. Connection p28 P99-110-4630

Z-2488-2021

July 30, 2021

Class II

Product summary

Firm
Paragon 28, Inc.
Event
Event 88553
Status
Terminated
Classification
Class II
Quantity
14 drill bits
Official record key
device-enforcement:Z-2488-2021

Official wording

Reason: A thin wall condition between the inner cannulation and the flutes of the drill which potentially could cause a fracture through the wall of the drill to the cannulation beneath.

Code information: Model Number: P99-110-4630 UDI Code: (01)00889795056849(10)HN120550 Lot Number: HN120550

Distribution pattern: U.S. Nationwide distribution in the states of CA, FL, GA, ID, KS, KY, LA, MI, MO, NC, NY, TX, VA, WA and WV. O.U.S.: N/A