Device product
Olympus URETERO-RENO FIBERSCOPE Model: URF-P6R
Z-1264-2022
Product summary
- Event
- Event 90195
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 4 units
- Official record key
device-enforcement:Z-1264-2022
Official wording
Reason: Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.
Code information: UDI-DI: 04953170340833 Serial Number: 2836209F 2835978 2837272 2836299
Distribution pattern: CA FL MA MD MI NJ OH PR WI
Derived failure modes
-
Unknown
Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.