VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), Catalog Number 6802445, IVD --- Ortho Clinical Diagnostics. The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods).
Z-0968-2015
Product summary
- Event
- Event 69830
- Status
- Terminated
- Classification
- Class II
- Quantity
- 339 Total: USA - 102, Foreign - 237
- Official record key
device-enforcement:Z-0968-2015
Official wording
Reason: Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
Code information: Software Version 3.1 & Below; Serial Numbers J46000108 - J46000492 (the serial numbers are sequential; J Numbers are analogous to serial numbers)
Distribution pattern: Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Derived failure modes
-
Device software or design
Software Anomaly: the firm has identified an anomaly