Device product
ARCHITECT SHBG Reagent The ARCHITECT SHBG assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sex hormone binding globulin (SHBG) in human serum and plasma on the ARCHITECT iSystem. The ARCHITECT SHBG assay is used as an aid in the diagnosis of androgen disorders.
Z-2613-2017
Product summary
- Event
- Event 77070
- Status
- Terminated
- Classification
- Class II
- Quantity
- 7981 kits
- Official record key
device-enforcement:Z-2613-2017
Official wording
Reason: Firm has identified that ARCHITECT SHBG results generated with the Potassium EDTA specimen tube type may demonstrate a negative shift relative to results generated with serum. There is a potential for falsely decreased ARCHITECT SHBG results to be generated when using the ARCHITECT SHBG assay with the Potassium EDTA specimen tube type.
Code information: List number: 8K26-20, Lot Numbers: 01916E000, 00316F000, 01816G000, 00916I000, 07316I000, 00916L000 ; List Number: 8K26-25, Lot Numbers: 01816E000, 00216F000, 01716G000, 00816I000, 07216I000
Distribution pattern: Worldwide Distribution - USA (nationwide) and to the countries of : ARGENTINA, ARMENIA, AUSTRIA, AZERBAIJAN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BRUNEI, CANADA, CHINA, CHILE, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DEM. REP. OF THE CON, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, HUNGARY, INDIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JORDAN, KAZAKHSTAN, KUWAIT, LATVIA, LEBANON, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MONTENEGRO, NAMIBIA, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PARAGUAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UAE, UNITED KINGDOM, URUGUAY, VIETNAM
Derived failure modes
-
Unknown
Firm has identified that ARCHITECT SHBG results generated with the Potassium EDTA specimen tube type may demonstrate a negative shift relative to results generated with serum. There is a potential for falsely decreased ARCHITECT SHBG results to be generated when using the ARCHITECT SHBG assay with the Potassium EDTA specimen tube type.