Skip to content
Recall Observatory FDA recall evidence

Device product

SureSigns VS3 NBP, SpO2, Product Number 863071. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).

Z-1319-2019

April 19, 2019

Class II

Product summary

Firm
Philips North America, LLC
Event
Event 82684
Status
Ongoing
Classification
Class II
Quantity
12543
Official record key
device-enforcement:Z-1319-2019

Official wording

Reason: System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.

Code information: Serial Numbers: All Including UDIs: (01)00884838028890(21) and (01)00884838087075(21) followed by serial number.

Distribution pattern: Worldwide distribution. US Nationwide including Puerto Rico, Andorra, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Dominican Republic, Egypt, France, French Guiana, French Polynesia, Germany, Gibraltar, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea Republic, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Martinique, Mexico, Morocco, Mozambique, Myanmar, Namibia, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Philippines, Portugal, Qatar, Reunion, Romania, Rwanda, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Kingdom, United Arab Emirates,Venezuela, Vietnam, and Zimbabwe.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.