Skip to content
Recall Observatory FDA recall evidence

Device product

Smith & Nephew 5.5 mm Dyonics Bonecutter ,Electroblade Full Radius Part Number: 7205962 Product Usage: The Dyonics Electroblade Resector is indicated for use in arthroscopic surgical procedures of large and small articular cavities. The Dyonics Electroblade Resector is effective in soft and osseous tissue resection and in hemostasis of bleeding vessels. It is intended for arthroscopic procedures using saline solution, Ringers lactate or other conductive solutions as an irrigant under direct or video assisted fiberoptic visualization.

Z-2145-2012

July 14, 2012

Class II

Product summary

Firm
Smith & Nephew, Inc. Endoscopy Division
Event
Event 62617
Status
Terminated
Classification
Class II
Quantity
108 units
Official record key
device-enforcement:Z-2145-2012

Official wording

Reason: Device labeled with an expiry period of 48 months instead of 36 months; an additional 12 months for which product and package reliability has not been validated.

Code information: Lot Numbers: 655844R, 950143R, 950144R, 950145R, 950146R

Distribution pattern: Worldwide Distribution - US Nationwide and the countries of : Australia, Canada, Korea, Turkey, UAE, and Japan.