Device product
Target Detachable Coils: TARGET HELICAL NANO 2.5MM X 4CM MODEL Number:M0035452540 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Z-2183-2016
Product summary
- Event
- Event 74425
- Status
- Terminated
- Classification
- Class II
- Quantity
- 45,302 (8,188 in the U.S.) - total, all model numbers
- Official record key
device-enforcement:Z-2183-2016
Official wording
Reason: Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
Code information: MODEL Number:M0035452540; Lot Number/Expiration Date: 18440351 30-Jun-18 18539018 31-Aug-18 18439805 30-Jun-18 18519562 31-Jul-18 18468754 30-Jun-18 18922401 31-Mar-19 19060307 31-May-19 18970593 30-Apr-19 18898116 28-Feb-19 19034354 31-May-19 18875384 28-Feb-19.
Distribution pattern: Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.
Derived failure modes
-
Unknown
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.