Carestream Touch Prime, Catalog # 1738830, and Touch Prime XE, Catalog # 1738822, Ultrasound System Diagnostic ultrasound imaging or fluid flow analysis of the human body

Z-1052-2017

November 21, 2016

Class II

Product summary

Firm
Carestream Health Inc
Event
Event 76213
Status
Terminated
Classification
Class II
Quantity
US: 14 units; Foreign: 16 units
Official record key
device-enforcement:Z-1052-2017

Official wording

Reason: Software error; Carestream Health Inc, received a complaint stating that when a user accidentally obtains a measurement value of 0 and corrects the value in the report, the resulting measurement unit is not displayed, i.e., centimeters or millimeters. As such, the user expects that the measurement is taken calculated in centimeters, consistent with other values in the report. In actuality, the measurement is taken in millimeters. When this updated measurement is used in an average calculation, the result appears incorrect as two measurements are interpreted as centimeters while the user corrected value is interpreted as millimeters. If the user selects a Calc Result display as Min or Max, the values are also interpreted as millimeters when centimeters were expected.

Code information: Serial Numbers: US units: 106, 112, 114, 116, 1021, 1037, 1038, 1044, 1057, 1067, 1072, 1074, 1079, 1090; OUS units: 1005, 1006, 1009, 1011, 1015, 1017, 1018, 1020, 1031, 1041, 1042, 1043, 1045, 1081, 1086, 1089.

Distribution pattern: Worldwide Distribution - US including SC, GA, TX, OH, IA, NY and foreign distribution to Switzerland and Malaysia

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software error