Skip to content
Recall Observatory FDA recall evidence

Device product

Sunquest Laboratory SMART & Sunquest Laboratory SMART Select

Z-2216-2012

April 28, 2011

Class III

Product summary

Firm
Sunquest Information Systems, Inc.
Event
Event 62344
Status
Terminated
Classification
Class III
Quantity
103 sites
Official record key
device-enforcement:Z-2216-2012

Official wording

Reason: In Online Entry (OEx) there are two scenarios where a Test result from one container may file to a Test on another container. In order for this to occur: 1) Site parameter SIT 1,2,1,26 (3,217) [Restrict resulting of tests to owner container (<Y>/N)]=Y. 2) The Tests are ordered on the same accession number and 3) The Test codes are defined to have the same upload code as either replicate or

Code information: Version 6.3.0 and Later

Distribution pattern: Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    In Online Entry (OEx) there are two scenarios where a Test result from one container may file to a Test on another container. In order for this to occur: 1) Site parameter SIT 1,2,1,26 (3,217) [Restrict resulting of tests to owner container (<Y>/N)]=Y. 2) The Tests are ordered on the same accession number and 3) The Test codes are defined to have the same upload code as either replicate or