Skip to content
Recall Observatory FDA recall evidence

Device product

CPAP Mask Cushion, Model Nos. PB781S, PB781M, PB781L, 781S, 781M and 781L Product Usage: A CPAP mask is attached to an end user s face and a CPAP machine, to treat end users with sleep apnea. The replacement cushion is a non-durable component that requires replacement every 3 - 6 months. The cushion is used for both comfort and to provide a seal between the face and the mask.

Z-2741-2018

June 18, 2018

Class I

Product summary

Firm
Compass Health Brands (Corporate Office)
Event
Event 80545
Status
Terminated
Classification
Class I
Quantity
742
Official record key
device-enforcement:Z-2741-2018

Official wording

Reason: A May 2015 design change in the cushion seal replacement part and accompanying elbow replacement part resulted in incompatibility of the new cushion seal with the previous design of the elbow. Use of the new cushion seal with the previous design elbow could lead to unacceptable CO2 porting. As there was no model number change for this design change, customers may not be aware of the incompatibility.

Code information: All units manufactured after May 1, 2015

Distribution pattern: US nationwide distribution to AL, CT, FL, GA, KY, MS, NC, NY, TN, and TX.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    design change, customers may not be aware of the incompatib