Skip to content
Recall Observatory FDA recall evidence

Device product

Vital-Port Vascular Access System, Subcutaneous, implanted, intravascular infusion port and catheter, IP, IP-S, Product usage: The devices are intended for use in patient therapy requiring long-term vascular access for infusion therapy and/or blood sampling.

Z-1223-2018

November 28, 2017

Class II

Product summary

Firm
Cook Vascular Inc.
Event
Event 79425
Status
Terminated
Classification
Class II
Quantity
31,006 Units
Official record key
device-enforcement:Z-1223-2018

Official wording

Reason: Non-coring needle provided with the Cook Vital-Port¿ Vascular Assess System (Vital-Port) may cut or dislodge a core or sliver of material from the Vital-Port septum when the non-coring needle is inserted into the Vital-Port. This needle is used on the initial implant of the Vital-Port. Vital-Port products that have been successfully placed in patients are not impacted by this recall.

Code information: IPI, P-S

Distribution pattern: The devices have been distributed within the United States and the following countries: United Arab Emirates, Argentina, Austria, Australia, Barbados, Belgium, Bulgaria, Bermuda, Brazil, Canada, Switzerland, Colombia, Costa Rica, Cyprus, Czech Republic, Germany, Denmark, Spain, Finland, Faroe Islands, France, United Kingdom, Guatemala, Hong Kong, Hungary, Indonesia, Ireland, Iceland, Italy, Jamaica, Jordan, Japan, Kenya, South Korea, Kuwait, Lebanon, Latvia, Morocco, Macedonia, Mexico, Netherlands, New Zealand, Oman, Panama, Poland, Portugal, Serbia, Saudi Arabia, Sweden, San Marino, El Salvador, Turkey, Trinidad & Tobago, Ukraine, Uruguay, Venezuela, Virgin Islands, Kosovo, China, and South Africa.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Non-coring needle provided with the Cook Vital-Port¿ Vascular Assess System (Vital-Port) may cut or dislodge a core or sliver of material from the Vital-Port septum when the non-coring needle is inserted into the Vital-Port. This needle is used on the initial implant of the Vital-Port. Vital-Port products that have been successfully placed in patients are not impacted by this recall.