Skip to content
Recall Observatory FDA recall evidence

Device product

Break-Away Femoral Nozzle, Product Number: 0206-512-000, UDI: (01)4546540039415 for orthopedic use.

Z-1243-2018

November 09, 2017

Class II

Product summary

Firm
Stryker Instruments Div. of Stryker Corporation
Event
Event 79418
Status
Terminated
Classification
Class II
Quantity
384
Official record key
device-enforcement:Z-1243-2018

Official wording

Reason: Lack of sterility assurance as bioburden levels were found to be higher than internal acceptable rates.

Code information: Lot Number: 17279012 UDI: (01)4546540039415

Distribution pattern: Distributed in all 50 U.S. States and the District of Columbia. Distributed in Australia, Argentina, Japan, Canada, Chile, Korea, Spain, France, Italy, Netherlands, and United Kingdom.

Derived failure modes

  • Microbial contamination

    reason.microbial_contamination · v1.0.0

    bioburden levels were found
  • Sterility assurance

    reason.sterility_assurance · v1.0.0

    Lack of sterility